ASSESSMENT OF SETUP ERRORS IN INTENSITY-MODULATED RADIATION THERAPY TECHNIQUE USING IVIEWGT SYSTEM IN RADIATION THERAPY DEPARTMENT - MILITARY HOSPITAL 175


Authors

  • Ngọc Nguyễn Thế Bệnh viện Quân y 175
  • Hà Nguyễn Văn Bệnh viện Quân y 175
  • Luân Phạm Thành Bệnh viện Quân y 175
  • Minh Phạm Văn Bệnh viện Quân y 175
  • Hưng Đỗ Thị Bệnh viện Trung ương Quân đội 108
DOI: https://doi.org/10.59354/ydth175.2024.294

Keywords:

Radiation therapy, uncertainty in radiation therapy, systematic errors, random errors

Abstract

Aim: The study is for assessment of errors in three dimensions of patient setup before irradiation delivery and calculating setup margins for head and neck cases with intensity modulated radiation therapy technique.

Methode: 937 orthogonal and corresponding portal iViewGT images of 51 patients with IMRT head and neck treatment were compared with digitally reconstructed radiographs (DRR) from computed tomography simulation scans. The displacements for vertical (A.P), lateral (L.R), longitudinal (S.I) directions were collected. The systematic errors () and random errors () of radiation unit were calculated. Setup margins (A) for 3 dimentional directions were caculated for target volume in head and neck region.

Result: The displacements in vertical (A.P), lateral (L.R) and longitudinal (S.I) directions were 2.2 ± 2.4 mm, -0.03 ± 1.7 mm, 1.95 ± 2.1 mm respectively. The systematic error for each directions were:  = 1.15 mm,  = 0.74 mm,  = 1.17 mm, the random errors for each directions were:  = 2.1 mm,  = 1.5 mm ,  = 1.7 mm. Setup margins for treatment volume in head and neck region in each directions were: = 4.3 mm,  = 2.9 mm, = 4.1 mm.

Conclusion: Setup errors of patients with head and neck IMRT treatments in Radiation therapy department is in the tolerance for IMRT treatment, well compares with published studies. The setup margin was less than 5 mm in all dimentions and was less than 3 mm if patients have more than 2 times a week iViewGT checks before having treatment. Portal images iViewGT checking is a solution for safe and presice patient setup before delivering IMRT plan.

References

Global Cancer Statistics 2020: “GLOBOCAN Estimates of Incidence and Mortality Worldwide for 36 Cancers in 185 Countries”.

Marcel van Herk (2004). “Errors and Margins in Radiotherapy”. Seminars in Radiation Oncology, Vol 14, No 1 (January), 2004:p.52-64.

ICRU 29/50/62/83 (International Commission on Radiation Units & Measurements) – IAEA

Matteo Tamponi et al. (2013). “Random and systematic set-up errors in threedimensional conformal radiotherapy – impact on planning target volume margins: the experience of the Radiation Oncology Centre of Sassari”. Journal of Radiotherapy in Practice (2014). p.169-178.

Tejpal Gupta et al. (2007). “Assessment of three-dimensional set-up errors in conventional head and neck radiotherapy using electronic portal imaging device”. p.4-8.

Jiqing su et al. “Different setup errors assessed by weekly cone-beam computed tomography on different registration in nasopharyngeal carcinoma treated with intensitymodulated radiation therapy”. Journal of OncoTargets and Therapy (2015).p.2545-2552.

Valeria Mongioj et al. “Set-up errors analyses in IMRT treatments for nasopharyngeal carcinoma to evaluate time trends, PTV and PRV margins” Journal of Acta Oncologica, (2011).p.61–71.

Durim Delishaj et al. “Set-up errors in head and neck cancer treated with IMRT technique assessed by cone-beam computed tomography: a feasible protocol” Radiation Oncology Journal (2018).p.54-62.

Hunt MA, Kutcher GJ, Burman C, et al: The effect of setup uncertainties on the treatment of nasopharynx cancer. Int J Radiat Oncol Biol Phys 27:437-447, 1993

Urie MM, Goitein M, Doppke K, et al: The role of uncertainty analysis in treatment planning. Int J Radiat Oncol Biol Phys 21:91-107, 1991

Hurkmans CW, Remeijer P, Lebesque JV, et al: Set-up verification using portal imaging; review of current clinical practice. Radiother Oncol 58:105-120, 2001.

Abstract View: 51
PDF Downloaded: 41

Published

30-09-2024

How to Cite

Nguyễn Thế , N., Nguyễn Văn , H., Phạm Thành , L., Phạm Văn , M., & Đỗ Thị , H. (2024). ASSESSMENT OF SETUP ERRORS IN INTENSITY-MODULATED RADIATION THERAPY TECHNIQUE USING IVIEWGT SYSTEM IN RADIATION THERAPY DEPARTMENT - MILITARY HOSPITAL 175. Journal of 175 Practical Medicine and Pharmacy, (39), 13. https://doi.org/10.59354/ydth175.2024.294